In the field of reliability environmental testing equipment, the compressor serves as the core refrigeration component within a constant temperature and humidity test chamber. It plays a critical role in determining the overall performance and efficiency of the system. Yashilin Instruments, combining extensive research, development, and production experience, has partnered with renowned brands such as Taikang from France and Zell from Germany to enhance the quality and durability of their constant temperature and humidity test chambers. Based on this collaboration, we have compiled the following maintenance guidelines for the Taikang Bitzer compressor:
First, it is essential to ensure that the environment where the constant temperature and humidity test chamber is installed meets specific requirements:
1. The floor should be level, well-ventilated, and free from flammable, explosive, or corrosive gases, as well as dust.
2. There should be no strong electromagnetic radiation sources nearby.
3. Sufficient space must be left around the equipment for easy access and maintenance.
4. The ambient temperature should be relatively low to prevent overheating.
5. The relative humidity should not exceed 85% RH, especially during summer when air conditioning is in use.
6. The atmospheric pressure should be between 86 kPa and 106 kPa.
Second, maintaining a clean and well-ventilated working environment is crucial. Dust accumulation can significantly affect the performance and lifespan of the compressor. Always keep the surface of the compressor clean. If the test chamber is not in use for an extended period, store it in a dust-free area to prevent damage.
Third, the power supply voltage provided to the test chamber should remain stable. Fluctuations in voltage can lead to excessive current flow, which may cause serious damage to the compressor over time. It is highly recommended to use a voltage stabilizer, especially for high and low temperature test chambers or constant temperature and humidity test chambers.
Fourth, avoid frequent starting of the compressor. A minimum interval of 15 minutes between startups is necessary to protect the compressor from unnecessary wear and tear. Compressor maintenance is a key part of daily upkeep, as the compressor is essentially the "heart" of the constant temperature and humidity test chamber.
Yashilin Instruments offers a wide range of testing equipment, including low-temperature test chambers, constant temperature and humidity test chambers, high-low temperature test chambers, thermal shock test chambers, UV aging test chambers, xenon lamp aging test chambers, rain test chambers, high-low temperature cyclic test chambers, salt fog testing machines, vibration tables, and electronic product safety testing equipment. With a focus on quality and reliability, Yashilin continues to provide advanced solutions for industrial and laboratory testing needs.