**Jane Hydraulic Universal Testing Machine Maintenance Guide**
Maintaining a hydraulic universal testing machine is essential to ensure its accuracy, efficiency, and longevity. Here are some key maintenance tips for different parts of the machine.
**Machine Host Maintenance:**
1. Always apply anti-rust oil to the fixtures on the testing machine to prevent corrosion.
2. The jaws of the hydraulic universal testing machine are frequently used and can wear out over time. If they become too worn, it may lead to damage in the small pistons. To avoid this, clean the jaws regularly after each test and apply lubricating oil.
3. Clean and lubricate the sliding surfaces of the embedded steel plates, lining plates, and dovetail grooves with a thin layer of MoSâ‚‚ (molybdenum disulfide) grease to reduce friction.
4. Regularly check the screws on the jaws. If any are loose, tighten them immediately to maintain stability.
5. Inspect the sprocket transmission system periodically. If the chain becomes loose, adjust the tensioner to restore proper function.
**Hydraulic Oil Source Maintenance:**
1. Check both the host and oil source regularly for signs of leakage. If you notice any, replace the seals or gaskets as needed.
2. Based on the machine’s usage and oil condition, replace the oil filter and filter elements on schedule, and change the hydraulic oil to maintain performance.
3. When the machine is not in use for an extended period, turn off the power. If the machine is in standby mode, make sure the transfer switch is set to “load†rather than “fast retreat.†Leaving it in the “fast retreat†position could cause the electromagnetic reversing valve to remain energized, which may shorten the device's lifespan.
**Control System Maintenance:**
1. Periodically check the rear panel of the control unit to ensure all connections are secure. If any connections are loose, tighten them promptly.
2. After completing an experiment, it’s best to disconnect the controller from the computer if it won’t be used for a while. This helps protect the system from potential electrical issues.
3. Ensure that the interfaces on the controller match correctly. Incorrect connections can lead to damage to the equipment.
Proper maintenance not only improves the performance of your hydraulic universal testing machine but also extends its service life, ensuring reliable results for every test.